Implanted Dosimeters Identify Radiation Overdoses During IMRT for Prostate Cancer - Abstract

PURPOSE: Image-guided dose-escalated radiotherapy is the standard of care for the treatment of prostate cancer. Although many published methods are available that account for prostate motion during delivery, evidence demonstrating that the planned dose is actually delivered on a daily basis is lacking. We report our initial clinical experience using implantable dosimeters to quantify and adjust the dose received during intensity-modulated radiotherapy (IMRT).

METHODS AND MATERIALS: A total of 20 patients undergoing IMRT with cone-beam computed tomography (CT) image guidance for prostate cancer had the dose verification system with radiopaque metal-oxide-semiconductor field effect transistor dosimeters implanted before treatment planning. All patients underwent planning with CT simulation in the supine position with custom immobilization, and the implanted dosimeters were located in the IMRT plans. The predicted dose for each dosimeter was defined and compared with the wireless readings before and after each treatment session. Investigations by physicians and medical physicists were initiated for two or more discrepancies >6% for any five consecutive fractions or for any discrepancy ≥10%.

RESULTS: Using implanted in vivo dosimeters, dose measurements consistently >6% greater than the predicted values were observed during treatment for 3 of 20 prostate cancer patients who received IMRT with daily image guidance. A review of the daily cone-beam CT images revealed acceptable alignment of the prostate target volumes and implanted dosimeters but identified significant anatomic changes within the treated region. Repeat CT simulation and RT planning was performed, with resolution of the dose discrepancies in all 3 cases with the adoption of a new IMRT plan.

CONCLUSIONS: Our report illustrates the potential effect of implanted in vivo dosimetry for prostate IMRT and emphasizes the importance of careful planning and delivery with attention to systematic shifts or anatomic changes that could alter the dose distributions.

Written by:
Den RB, Nowak K, Buzurovic I, Cao J, Harrison AS, Lawrence YR, Dicker AP, Showalter TN   Are you the author?
Department of Radiation Oncology, Kimmel Cancer Center, Jefferson Medical College of Thomas Jefferson University, Philadelphia, Pennsylvania

Reference: Int J Radiat Oncol Biol Phys. 2012 Jul 1;83(3):e371-6.
doi: 10.1016/j.ijrobp.2011.12.094

PubMed Abstract
PMID: 22633553