There are currently limitations associated with the prostate biopsy procedure, which is the most commonly used method for a definitive diagnosis of prostate cancer. With the use of two-dimensional (2D) transrectal ultrasound (TRUS) for needle-guidance in this procedure, the physician has restricted anatomical reference points for guiding the needle to target sites. Further, any motion of the physician's hand during the procedure may cause the prostate to move or deform to a prohibitive extent. These variations make it difficult to establish a consistent reference frame for guiding a needle. We have developed a 3D navigation system for prostate biopsy, which addresses these shortcomings. This system is composed of a 3D US imaging subsystem and a passive mechanical arm to minimize prostate motion. To validate our prototype, a series of experiments were performed on prostate phantoms. The 3D scan of the string phantom produced minimal geometric distortions, and the geometric error of the 3D imaging subsystem was 0.37 mm. The accuracy of 3D prostate segmentation was determined by comparing the known volume in a certified phantom to a reconstructed volume generated by our system and was shown to estimate the volume with less then 5% error. Biopsy needle guidance accuracy tests in agar prostate phantoms showed that the mean error was 2.1 mm and the 3D location of the biopsy core was recorded with a mean error of 1.8 mm. In this paper, we describe the mechanical design and validation of the prototype system using an in vitro prostate phantom. Preliminary results from an ongoing clinical trial show that prostate motion is small with an in-plane displacement of less than 1 mm during the biopsy procedure.

Written by:
Bax J, Cool D, Gardi L, Knight K, Smith D, Montreuil J, Sherebrin S, Romagnoli C, Fenster A   Are you the author?

Reference:
Med Phys. 2008 Dec;35(12):5397-410

PubMed Abstract
PMID: 19175099

UroToday.com Prostate Cancer Section

Reader Comments

Please log-in or register in order to submit comments.

Powered by AkoComment!

 
User Rating: / 0
PoorBest


 

Bookmark and Share
< Prev   Next >

Member's Section

Login

Sign Up

Quick Search

Meet the Expert


All Experts



Featured Conference

Media and Publisher

Advertising Rates

Working with Industry

Case Studies
Sponsorship Opportunities

Prostate Cancer
Sponsored by